China NMPA vs Malaysia MDA — Recognition of Foreign Pre-Market Approval

/
/
China NMPA vs Malaysia MDA — Recognition of Foreign Pre-Market Approval
China NMPA vs Malaysia MDA medical device pre-market approval recognition

China NMPA vs Malaysia MDA — Recognition of Foreign Pre-Market Approval

Recognition of foreign pre-market approvals is a clear and growing global trend in medical-device regulation. The terminology varies—regulatory reliance, abridged review, verification, recognition of overseas assessments, etc.—but the underlying principle is the same:

Regulators increasingly leverage the work already performed by trusted foreign regulators instead of repeating the entire assessment from scratch.

Historically, a manufacturer entering each country generally had to undergo a separate national regulatory assessment:

Foreign approval → New national assessment → New national approval

The emerging model is:

Trusted foreign approval → Reliance / verification / abridged review → Faster national market access

This is being driven by the need to reduce regulatory duplication, shorten time-to-market, conserve regulatory resources and improve patient access. However, this concept has yet to be fully accepted by many authorities.

The following comparison highlights a key difference between China NMPA and Malaysia MDA in their treatment of foreign pre-market approvals.

Aspect China NMPA Malaysia MDA
Accepts foreign pre-market approval? Yes, as supporting evidence Yes, for recognised foreign authorities under the verification pathway
Does foreign approval replace local approval? No No
Can foreign approval reduce local assessment? No Yes
Formal verification/reliance pathway? No Yes
Foreign authorities recognised? Foreign competent-authority marketing certificate is required for imported devices, as supporting evidence Yes. MDA maintains a defined list of recognised competent authorities
Use of foreign technical/clinical evidence Yes, subject to NMPA requirements Yes, through verification of evidence
Local conformity assessment repeated? Generally, NMPA performs its own regulatory review Full assessment can be avoided/reduced if eligible for verification
Local product registration still required? Yes Yes

While NMPA accepts foreign regulatory approvals as supporting evidence for imported medical devices, manufacturers must generally undergo NMPA’s own regulatory review and foreign approval does not reduce the local assessment in a broad reliance pathway. In contrast, MDA has established a formal verification pathway that allows eligible medical devices with approvals from recognised foreign regulatory authorities to leverage existing conformity-assessment evidence, potentially reducing duplication of assessment, regulatory time and cost. The eligibility criteria for accessing Malaysia’s Conformity Assessment by Way of Verification Route are set out in the Medical Device Authority (MDA) Guidance Document.

In both markets, foreign approval does not replace local market authorisation; NMPA or MDA approval remains necessary before the device can be placed on the respective market.

Medical device pre-market approval can be a lengthy and costly process. By recognising eligible foreign pre-market approvals, Malaysia MDA has significantly accelerated market entry and reduced regulatory costs, making Malaysia a more attractive gateway for medical device manufacturers seeking access to the Malaysian and regional market.

The following table provides a comparison of the typical market-entry turnaround time and estimated regulatory costs for medical device registration under China NMPA and Malaysia MDA.

China NMPA Malaysia MDA
TAT CAB conformity assessment Not applicable Class B–D: 30 - 60 working days
after complete submission
Regulatory evaluation Class I: 5 working days
Class II: 60 working days
Class III: 90 working days
Class A–D: 30 - 60 working days
after complete submission
Typical overall time Class II: ~12–24+ months
Class III: ~18–36+ months
Class B, C & D: ~3–5+ months
Cost Regulatory evaluation Class I: Not required
Class II: RMB 210,900 ≈ US$29,500
Class III: RMB 308,800 ≈ US$43,200
Class A: RM1,250 ≈ US$315
Class B: RM1,250 ≈ US$315
Class C: RM2,500 ≈ US$625
Class D: RM3,750 ≈ US$940
CAB full conformity assessment (FCA) Not applicable Class B: ~ US$3,750
Class C: ~ US$6,250
Class D: ~ US$8,750
CAB conformity assessment via verification route (VR) Not applicable Class B, C & D: ~ US$300
Estimated total market-entry cost (excluding profession regulatory affairs supports fee) Class II: ~US$30,000
Class III: ~US$50,000
Class A: ~ US$315
Class B: ~ US$4,500 (FCA)/650 (VR)
Class C: ~ US$6,500 (FCA)/ 950 (VR)
Class D: ~ US$10,000 (FCA)/1,250 (VR)
Relative cost Higher Lower
Relative speed Slower Faster

In our blog, “Which Southeast Asian Country Should You Enter First?”, Malaysia is identified as a “practical regional entry point” for foreign medical device manufacturers looking to test the market, gain ASEAN experience, and establish a foundation for regional expansion.

Find the Right Regulatory Pathway into Malaysia

If your medical devices have already obtained pre-market clearance from an MDA-recognised reference authority, we invite you to contact Global Rep to assess your regulatory readiness and identify the most efficient pathway to enter the Malaysian and wider regional markets.

With an established market portfolio and extensive experience supporting Chinese medical device manufacturers, Global Rep understands the regulatory and commercial challenges of expanding into Malaysia and ASEAN. Our practical market experience, together with positive testimonials from Chinese manufacturers we have supported, demonstrates our commitment to delivering reliable, responsive and cost-effective regulatory solutions.

Recent Posts

China Malaysia medical device regulatory bridge for market access

Your China–Malaysia Medical Device Regulatory...

For a Chinese medical device manufacturer planning to enter Malaysia,…
China NMPA vs Malaysia MDA medical device pre-market approval recognition

China NMPA vs Malaysia MDA...

Recognition of foreign pre-market approvals is a clear and growing…
Southeast Asian medical device markets and regulatory pathways

Which Southeast Asian Country Should...

For a medical device manufacturer planning to expand into Southeast…